Methods Section of the Paper

 

Please use the charts below to create the configuration specific to your application. Following is an exampleof a typical description. Please contact the lab personnel if you wish assistance with the wording:

Two-color flow cyotmetric analysis was performed using a FACScan flow cytometer (BDBiosciences, San Jose, CA) equipped with an air-cooled 15mW argon ion laser tuned to 488nm. The emission fluorescence of (yada, yada conjugated to IgG…blah blah..) was detected and recorded through a 530/30 bandpass filter in the FL1 channel. (Something red) was detected in the FL2 channel through a 585/42 bandpass filter. List mode data files consisting of 10,000 events gated on FSC (forward scatter) vs SSC (side scatter) were acquired and analyzed using CellQuest PRO software (BD Biosciences, San Jose, CA) at a rate of 200-400 events per second. Appropriate electronic compensation was adjusted by acquiring cell populations stained with each dye/fluorophore individually, as well as an unstained control.

BD Biosciences FACSAria Configuration

Excitation

Laser

Emission

Detection Filter

Detector

Example

(there are many others)

488 nm (blue)

1.5 ND Filter

FSC Diode

780/60 (750-810 nm)

PMT A

PE-Cy7

695/40 (675-715 nm)

PMT B

PerCP, PE- Cy5.5

610/20 (600-620 nm)

PMT C

PE-Texas Red

576/26 (563-589 nm)

PMT D

PE

530/30 (515-545 nm)

PMT E

FITC

488/10 (483-493 nm)

PMT F

SSC only

633 nm (red)

780/60 (750-810 nm)

PMT A

APC-Cy7

660/20 (650-670 nm)

PMT B

APC

407 nm (violet)

530/30 (515-545 nm)

PMT A

Alexa-430

450/40 (430-470 nm)

PMT B

DAPI

BD Biosciences FACScan Configuration

Excitation

Laser

Emmision

Detection Filter

Detector

Example

488 nm (blue)

1.0 ND Filter

FSC Diode

 

488/10 (483-493 nm)

SSC Detector

530/30 (515-545 nm)

FL1 Detector

FITC

585/42 (564-606 nm)

FL2 Detector

PE

650 LP (650-800 nm)

FL3 Detector

PerCP